What do the alarms and lights on the Respironics SimplyGo mean?

The Respironics SimplyGo shows alarms on its display screen with a sound, and has an alarm silence button on the control panel.

No Breath Detected: In pulse mode, if the unit doesn't detect a breath, it delivers oxygen at 12 breaths per minute. It alarms after 2 minutes with no breath detected and powers down after 30 minutes.
Battery Check: Press the white button on the battery and count the lit LED lights.
Screen: Shows the operating mode, setting, battery status, breath indicator, external power and alarm status.

For more information, visit Audible Alerts and Visible Signals for the Respironics SimplyGo.