What do the alarms and lights on the Inogen One Rove 4 mean?
The Inogen One Rove 4 uses a green light, a yellow light and beeps to show its status. A green light means a breath was detected. A yellow light means the unit needs your attention, and a flashing light is a higher priority than a solid one.
Level 1: One short beep
Level 2: One beep and a solid yellow light
Level 3: Two beeps and a solid yellow light
Level 4: Three beeps every 25 seconds and a flashing yellow light
A low priority alarm sounds when the battery has less than 10 minutes left, and it becomes a higher priority when the battery is empty. Plug the unit into AC or DC power, or turn it off and change the battery. The bell button turns the No Breath Detected beep on or off.
To see past alarms, plug the unit in, turn it off, and hold the + button for 5 seconds. The alarm log is also in the Inogen Connect app.
For more information, visit Audible Alerts and Visible Signals for the Inogen Rove 4.
Level 1: One short beep
Level 2: One beep and a solid yellow light
Level 3: Two beeps and a solid yellow light
Level 4: Three beeps every 25 seconds and a flashing yellow light
A low priority alarm sounds when the battery has less than 10 minutes left, and it becomes a higher priority when the battery is empty. Plug the unit into AC or DC power, or turn it off and change the battery. The bell button turns the No Breath Detected beep on or off.
To see past alarms, plug the unit in, turn it off, and hold the + button for 5 seconds. The alarm log is also in the Inogen Connect app.
For more information, visit Audible Alerts and Visible Signals for the Inogen Rove 4.