What do the alarms and lights on the Inogen One G5 mean?
The Inogen One G5 uses a green light, a yellow light and beep patterns to show its status. A green light means a breath was detected. A yellow light means the unit needs your attention, and a flashing yellow light is more urgent than a solid one.
Green Light: Breath detected, normal operation
Notification: One short beep with an icon on the screen
Low Priority: 2 beeps and a solid yellow light
Medium Priority: 3 beeps repeated every 25 seconds and a flashing yellow light
High Priority: 5 beeps repeated every 10 seconds and a flashing yellow light
A low priority alert sounds when the battery has about 10 minutes of power left. A medium priority alert sounds when no breath has been detected for 60 seconds, which can mean a problem with the cannula connection or placement. A high priority alert means the battery is empty, the unit is too hot or too cold, the battery is overheating, or several alerts are happening at once.
You can turn the breath detection beep on or off and set the alert volume from 1 to 4, but priority alerts can't be silenced.
For more information, visit Audible Alerts and Visible Signals for the Inogen One G5.
Green Light: Breath detected, normal operation
Notification: One short beep with an icon on the screen
Low Priority: 2 beeps and a solid yellow light
Medium Priority: 3 beeps repeated every 25 seconds and a flashing yellow light
High Priority: 5 beeps repeated every 10 seconds and a flashing yellow light
A low priority alert sounds when the battery has about 10 minutes of power left. A medium priority alert sounds when no breath has been detected for 60 seconds, which can mean a problem with the cannula connection or placement. A high priority alert means the battery is empty, the unit is too hot or too cold, the battery is overheating, or several alerts are happening at once.
You can turn the breath detection beep on or off and set the alert volume from 1 to 4, but priority alerts can't be silenced.
For more information, visit Audible Alerts and Visible Signals for the Inogen One G5.