The CAIRE SeQual Eclipse 5 uses green, yellow and red lights to show its status.
Green Light: Operating normally
Yellow Light: A non-critical warning or caution that needs your attention
Red Light: The unit isn't working properly and needs immediate attention, for example oxygen concentration below 70%
When you turn the unit on, it beeps and all the lights come on. The red light turns off once concentration rises above 70%, and the yellow light turns off once it reaches normal (above 85%). This can take up to five minutes.
To see an alarm code while the unit is alarming, press the No Smoking icon. If the Low Priority Technical Alarm light is on, you can keep using the unit and check the troubleshooting table in the user manual.
For more information, visit Audible Alerts and Visible Signals for the SeQual Eclipse 5.
Green Light: Operating normally
Yellow Light: A non-critical warning or caution that needs your attention
Red Light: The unit isn't working properly and needs immediate attention, for example oxygen concentration below 70%
When you turn the unit on, it beeps and all the lights come on. The red light turns off once concentration rises above 70%, and the yellow light turns off once it reaches normal (above 85%). This can take up to five minutes.
To see an alarm code while the unit is alarming, press the No Smoking icon. If the Low Priority Technical Alarm light is on, you can keep using the unit and check the troubleshooting table in the user manual.
For more information, visit Audible Alerts and Visible Signals for the SeQual Eclipse 5.